| Optics |
Double Beam (chopper mirror) |
| Monochromator |
aberration corrected Czerny-Turner monochromator |
| Holographic grating (1,600 lines/mm) |
| Wavelength range |
190-900 nm |
| Automated wavelength selection |
| Slit |
0.2 nm, 0.7 nm Manual setting |
| Background Correction |
D2-Lamp method |
| Lamp Turret |
2-lamps simultaneously lit (manual turret) |
| Lamp Mode |
Emission, Non-BGC, BGC-D2 |
| Frequency |
100Hz |
| Nebulizer |
Nebulizer integrated impact bead and jacket tip |
| Pt/Ir capillary |
| Chamber |
Polypropylene |
| Burner |
Fixed back/forward position and burner height |
| (Simply switching of Air/C2H2 and N2O/
C2H2 burner) |
| Titanium 10 cm slot burner (Optional high temperature burner) |
| Gas Control |
Manual setting of flow rate |
| Automatic Air/ N2O switching system |
| Safety |
Gas pressure monitoring to prevent flashback |
| Automatic flame monitoring |
| Safety interlock for misuse of burner |
| Automatic flame extinguish when power failure |
| Ignition |
Push ignite button |
| Software |
Software based on MS-Windows™ |
| Dimension and weight |
W690 x D425 x H370 mm, 38 kg |
| Power requirements |
AC220 V, 230 V, 50/60 Hz, 300 VA
(Certification of CE marking) |
| Ambient temperature and humidity range |
Temperature: 10~35 °C
Humidity: 45~80 % (but less than 70 % if temperature is
over 30 °C |
| Standard accessories |
Burner Head of 10 cm slot. Nebulizer assembly. Deuterium lamp.
Drain tube
PTFE tube for sample suction. Hose assembly for Air. Hose assembly for C2H2. Hose
band
RS-232C connection cable (9P-9P). AC cable for 220, 230 V |