Protein Sequencers
Protein sequencing via automated Edman degradation chemistry remains a simple and
robust solution for many labs, delivering unambiguous data that is much less complex
than the data obtained through mass spectrometry-based techniques. Utilizing state-of-the-art
HPLC analysis, high sensitivity detection and powerful new acquisition/analysis
software, Shimadzu’s PPSQ series is no exception, achieving greater simplicity and
reliability in the determination of amino acid sequences.
Features
Baseline Stability
Using the highly acclaimed Shimadzu HPLC, the PPSQ series protein sequencers separate
PTH-amino acids isocratically, improving baseline stability and allowing high-sensitivity
analysis of PTH-amino acids.
Superior Retention Time Reproducibility
Isocratic sequence analysis provides more stable retention times. That means peaks
detected in previous cycles can be cancelled using substation chromatogram processing,
making it easier for users to identify sequences.
Reduced Liquid Waste/Running Costs
Analysis in isocratic mode allows for eluents to be recycled, thereby reducing the
amount of mobile phase required along with reduced waste disposal, resulting in
significant cost savings. It also allows for quick instrument startup as the HPLC
can be left in a standby re-cycle mode.
Software
The PPSQ series’ specialized software incorporates control functions for the reaction unit and HPLC, enabling easier sequence analysis of proteins and peptides. Simple, easy-to-use data analysis functions simplify the following:
- Reprocessing of chromatograms
- Overlay of multiple chromatograms
- Automatic estimation of amino acid sequences
In addition, customized reports can be created, and yield rates can be displayed in graphical form with users able to freely select amino acids used for calculations.
Specifications
|
Main Unit
|
PPSQ-31A/33A
|
|
Reaction method
|
Edman degradation
|
|
Reaction time
|
PPSQ-31A: 47 min/cycle, PPSQ-33A: 48.5 min/cycle
|
|
Number of reactors
|
PPSQ-31A: 1, PPSQ-33A: 3
|
|
Sample immobilization method
|
Glass fiber disk (8 mm dia.) or PVDF membrane
|
|
Reactor temperature control
|
{Room temperature + 10 °C} to 60 °C
|
|
Converter temperature control
|
{Room temperature + 10 °C} to 70 °C
|
|
Column temperature control
|
{Room temperature + 10 °C} to 60 °C
|
|
Number of reagents/solvents
|
8 (one is an option)
|
|
Reagent/solvent delivery method
|
Nitrogen gas pressure
|
|
Dimensions
|
650 (W) x 520 (H) x 500 (D) mm
|
|
Weight
|
PPSQ-31A: 55 kg
|